Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes
PurposeTo compare the accuracy and reliability of intraocular lens (IOL) master and A-scan immersion biometry in silicone oil (SO)-filled eyes.MethodsA prospective, consecutive, nonrandomized study was performed in 34 SO-filled eyes of 34 patients, who underwent a pars plana vitrectomy, with SO remo...
Saved in:
Main Authors: | , , , |
---|---|
Format: | Journal |
Published: |
2018
|
Subjects: | |
Online Access: | https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84867511211&origin=inward http://cmuir.cmu.ac.th/jspui/handle/6653943832/51326 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Chiang Mai University |
id |
th-cmuir.6653943832-51326 |
---|---|
record_format |
dspace |
spelling |
th-cmuir.6653943832-513262018-09-04T06:12:23Z Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes P. Kunavisarut P. Poopattanakul C. Intarated K. Pathanapitoon Arts and Humanities Medicine PurposeTo compare the accuracy and reliability of intraocular lens (IOL) master and A-scan immersion biometry in silicone oil (SO)-filled eyes.MethodsA prospective, consecutive, nonrandomized study was performed in 34 SO-filled eyes of 34 patients, who underwent a pars plana vitrectomy, with SO removal and cataract surgery, as well as IOL implantation. Both IOL master and immersion A-scan were performed to measure the axial length (AXL) before SO removal. Three months after removal of the SO, AXL measurements using IOL master and refraction was performed. Accuracy of the two techniques was determined by a mean postoperative AXL using an IOL master and reliability was determined by mean actual postoperative refractive error.ResultsPreoperative mean AXL was 23.91±0.24 mm (range 21.33-28.61 mm) and 23.71±0.59 mm (range 19.27-36.18 mm) by IOL master and A-scan immersion, respectively. Postoperative mean AXL by IOL master was 23.90±0.23 mm (range 21.58-27.94 mm), which showed a statistically significant difference from the preoperative mean AXL by A-scan immersion (P0.005). The AXL measurement by IOL master also was more accurate than A-scan immersion by Pearsons correlation (0.966 vs 0.410). For reliability of the two techniques, the predictive postoperative refractive error in A-scan immersion (mean 1.79±1.04 D, range 14.62 to 16.41 D) was greater than that in IOL master (mean 0.60±0.23 D, range 2.74 to 2.33 D), with a statistically significant difference (P0.049).ConclusionIOL master had more accuracy and less deviation in predictive postoperative refractive error than A-scan immersion in SO-filled eyes. © 2012 Macmillan Publishers Limited All rights reserved. 2018-09-04T06:00:17Z 2018-09-04T06:00:17Z 2012-01-01 Journal 14765454 0950222X 2-s2.0-84867511211 10.1038/eye.2012.163 https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84867511211&origin=inward http://cmuir.cmu.ac.th/jspui/handle/6653943832/51326 |
institution |
Chiang Mai University |
building |
Chiang Mai University Library |
country |
Thailand |
collection |
CMU Intellectual Repository |
topic |
Arts and Humanities Medicine |
spellingShingle |
Arts and Humanities Medicine P. Kunavisarut P. Poopattanakul C. Intarated K. Pathanapitoon Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes |
description |
PurposeTo compare the accuracy and reliability of intraocular lens (IOL) master and A-scan immersion biometry in silicone oil (SO)-filled eyes.MethodsA prospective, consecutive, nonrandomized study was performed in 34 SO-filled eyes of 34 patients, who underwent a pars plana vitrectomy, with SO removal and cataract surgery, as well as IOL implantation. Both IOL master and immersion A-scan were performed to measure the axial length (AXL) before SO removal. Three months after removal of the SO, AXL measurements using IOL master and refraction was performed. Accuracy of the two techniques was determined by a mean postoperative AXL using an IOL master and reliability was determined by mean actual postoperative refractive error.ResultsPreoperative mean AXL was 23.91±0.24 mm (range 21.33-28.61 mm) and 23.71±0.59 mm (range 19.27-36.18 mm) by IOL master and A-scan immersion, respectively. Postoperative mean AXL by IOL master was 23.90±0.23 mm (range 21.58-27.94 mm), which showed a statistically significant difference from the preoperative mean AXL by A-scan immersion (P0.005). The AXL measurement by IOL master also was more accurate than A-scan immersion by Pearsons correlation (0.966 vs 0.410). For reliability of the two techniques, the predictive postoperative refractive error in A-scan immersion (mean 1.79±1.04 D, range 14.62 to 16.41 D) was greater than that in IOL master (mean 0.60±0.23 D, range 2.74 to 2.33 D), with a statistically significant difference (P0.049).ConclusionIOL master had more accuracy and less deviation in predictive postoperative refractive error than A-scan immersion in SO-filled eyes. © 2012 Macmillan Publishers Limited All rights reserved. |
format |
Journal |
author |
P. Kunavisarut P. Poopattanakul C. Intarated K. Pathanapitoon |
author_facet |
P. Kunavisarut P. Poopattanakul C. Intarated K. Pathanapitoon |
author_sort |
P. Kunavisarut |
title |
Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes |
title_short |
Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes |
title_full |
Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes |
title_fullStr |
Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes |
title_full_unstemmed |
Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes |
title_sort |
accuracy and reliability of iol master and a-scan immersion biometry in silicone oil-filled eyes |
publishDate |
2018 |
url |
https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84867511211&origin=inward http://cmuir.cmu.ac.th/jspui/handle/6653943832/51326 |
_version_ |
1681423749593694208 |