A stopped flow system with hydrodynamic injection for red blood cells osmotic fragility test: Possibility for automatic screening of beta-thalassemia trait
Simple instrumentation and procedure with stopped flow system coupled to hydrodynamic injection and spectrophotometric detection were developed for automatic osmotic fragility test (OFT). OFT is a test for abnormal red blood cells based on the kinetics of their rupturing in a hypotonic saline soluti...
Saved in:
Main Authors: | , , , , , |
---|---|
Format: | Journal |
Published: |
2018
|
Subjects: | |
Online Access: | https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=68249089807&origin=inward http://cmuir.cmu.ac.th/jspui/handle/6653943832/59478 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Chiang Mai University |
id |
th-cmuir.6653943832-59478 |
---|---|
record_format |
dspace |
spelling |
th-cmuir.6653943832-594782018-09-10T03:15:49Z A stopped flow system with hydrodynamic injection for red blood cells osmotic fragility test: Possibility for automatic screening of beta-thalassemia trait Supada Khonyoung Supaporn Kradtap Hartwell Jaroon Jakmunee Somchai Lapanantnoppakhun Torpong Sanguansermsri Kate Grudpan Chemistry Simple instrumentation and procedure with stopped flow system coupled to hydrodynamic injection and spectrophotometric detection were developed for automatic osmotic fragility test (OFT). OFT is a test for abnormal red blood cells based on the kinetics of their rupturing in a hypotonic saline solution. A portion of red cells was merged on-line with a hypotonic saline solution. They were mixed while flowing into the detector and were stopped for a short period for continuously monitoring the change of turbidity based on a transmission signal. In this work, a possible application of the system for screening of beta-thalassemia trait is demonstrated. Descriptions of instrumentation development and parameter optimization are presented. The system offers advantages over the conventional batch-wise OFT in terms of automation, precision, analysis time and sample volume. 2009 © The Japan Society for Analytical Chemistry. 2018-09-10T03:15:49Z 2018-09-10T03:15:49Z 2009-06-01 Journal 13482246 09106340 2-s2.0-68249089807 10.2116/analsci.25.819 https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=68249089807&origin=inward http://cmuir.cmu.ac.th/jspui/handle/6653943832/59478 |
institution |
Chiang Mai University |
building |
Chiang Mai University Library |
country |
Thailand |
collection |
CMU Intellectual Repository |
topic |
Chemistry |
spellingShingle |
Chemistry Supada Khonyoung Supaporn Kradtap Hartwell Jaroon Jakmunee Somchai Lapanantnoppakhun Torpong Sanguansermsri Kate Grudpan A stopped flow system with hydrodynamic injection for red blood cells osmotic fragility test: Possibility for automatic screening of beta-thalassemia trait |
description |
Simple instrumentation and procedure with stopped flow system coupled to hydrodynamic injection and spectrophotometric detection were developed for automatic osmotic fragility test (OFT). OFT is a test for abnormal red blood cells based on the kinetics of their rupturing in a hypotonic saline solution. A portion of red cells was merged on-line with a hypotonic saline solution. They were mixed while flowing into the detector and were stopped for a short period for continuously monitoring the change of turbidity based on a transmission signal. In this work, a possible application of the system for screening of beta-thalassemia trait is demonstrated. Descriptions of instrumentation development and parameter optimization are presented. The system offers advantages over the conventional batch-wise OFT in terms of automation, precision, analysis time and sample volume. 2009 © The Japan Society for Analytical Chemistry. |
format |
Journal |
author |
Supada Khonyoung Supaporn Kradtap Hartwell Jaroon Jakmunee Somchai Lapanantnoppakhun Torpong Sanguansermsri Kate Grudpan |
author_facet |
Supada Khonyoung Supaporn Kradtap Hartwell Jaroon Jakmunee Somchai Lapanantnoppakhun Torpong Sanguansermsri Kate Grudpan |
author_sort |
Supada Khonyoung |
title |
A stopped flow system with hydrodynamic injection for red blood cells osmotic fragility test: Possibility for automatic screening of beta-thalassemia trait |
title_short |
A stopped flow system with hydrodynamic injection for red blood cells osmotic fragility test: Possibility for automatic screening of beta-thalassemia trait |
title_full |
A stopped flow system with hydrodynamic injection for red blood cells osmotic fragility test: Possibility for automatic screening of beta-thalassemia trait |
title_fullStr |
A stopped flow system with hydrodynamic injection for red blood cells osmotic fragility test: Possibility for automatic screening of beta-thalassemia trait |
title_full_unstemmed |
A stopped flow system with hydrodynamic injection for red blood cells osmotic fragility test: Possibility for automatic screening of beta-thalassemia trait |
title_sort |
stopped flow system with hydrodynamic injection for red blood cells osmotic fragility test: possibility for automatic screening of beta-thalassemia trait |
publishDate |
2018 |
url |
https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=68249089807&origin=inward http://cmuir.cmu.ac.th/jspui/handle/6653943832/59478 |
_version_ |
1681425258746216448 |