Analytical performance of centrifuge-based device for clinical chemistry testing
Background: A centrifuge-based device has been introduced to the Samsung Blood Analyzer (SBA). The verification of this analyzer is essential to meet the ISO15189 standard. Methods: Analytical performance was evaluated according to the NCCLS EP05-A method. The results of plasma samples were compared...
Saved in:
Main Authors: | , |
---|---|
Other Authors: | |
Format: | Article |
Published: |
2018
|
Subjects: | |
Online Access: | https://repository.li.mahidol.ac.th/handle/123456789/13542 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Mahidol University |
id |
th-mahidol.13542 |
---|---|
record_format |
dspace |
spelling |
th-mahidol.135422018-06-11T11:32:23Z Analytical performance of centrifuge-based device for clinical chemistry testing Jamikorn Suk-Anake Chamras Promptmas Mahidol University Biochemistry, Genetics and Molecular Biology Background: A centrifuge-based device has been introduced to the Samsung Blood Analyzer (SBA). The verification of this analyzer is essential to meet the ISO15189 standard. Methods: Analytical performance was evaluated according to the NCCLS EP05-A method. The results of plasma samples were compared between the SBA and a Hitachi 917 analyzer according to the NCCLS EP09-A2-IR method. Percent recovery was determined via analysis of original control serum and spiked serum. Results: Within-run precision was found to be 0.00-6.61% and 0.96-5.99% in normal- and abnormal-level assays, respectively, while between-run precision was 1.31-9.09% and 0.89-6.92%, respectively. The correlation coefficients (r) were > 0.990. The SBA presented analytical accuracy at 96.64 ± 3.39% to 102.82 ± 2.75% and 98.31 ± 4.04% to 103.61 ± 8.28% recovery, respectively. Conclusions: The results obtained verify that all of the 13 tests performed using the SBA demonstrates good and reliable precision suitable for use in qualified clinical chemistry laboratory service. 2018-06-11T04:32:23Z 2018-06-11T04:32:23Z 2012-12-31 Article Clinical Laboratory. Vol.58, No.11-12 (2012), 1313-1318 10.7754/Clin.Lab.2012.120705 14336510 2-s2.0-84871568305 https://repository.li.mahidol.ac.th/handle/123456789/13542 Mahidol University SCOPUS https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84871568305&origin=inward |
institution |
Mahidol University |
building |
Mahidol University Library |
continent |
Asia |
country |
Thailand Thailand |
content_provider |
Mahidol University Library |
collection |
Mahidol University Institutional Repository |
topic |
Biochemistry, Genetics and Molecular Biology |
spellingShingle |
Biochemistry, Genetics and Molecular Biology Jamikorn Suk-Anake Chamras Promptmas Analytical performance of centrifuge-based device for clinical chemistry testing |
description |
Background: A centrifuge-based device has been introduced to the Samsung Blood Analyzer (SBA). The verification of this analyzer is essential to meet the ISO15189 standard. Methods: Analytical performance was evaluated according to the NCCLS EP05-A method. The results of plasma samples were compared between the SBA and a Hitachi 917 analyzer according to the NCCLS EP09-A2-IR method. Percent recovery was determined via analysis of original control serum and spiked serum. Results: Within-run precision was found to be 0.00-6.61% and 0.96-5.99% in normal- and abnormal-level assays, respectively, while between-run precision was 1.31-9.09% and 0.89-6.92%, respectively. The correlation coefficients (r) were > 0.990. The SBA presented analytical accuracy at 96.64 ± 3.39% to 102.82 ± 2.75% and 98.31 ± 4.04% to 103.61 ± 8.28% recovery, respectively. Conclusions: The results obtained verify that all of the 13 tests performed using the SBA demonstrates good and reliable precision suitable for use in qualified clinical chemistry laboratory service. |
author2 |
Mahidol University |
author_facet |
Mahidol University Jamikorn Suk-Anake Chamras Promptmas |
format |
Article |
author |
Jamikorn Suk-Anake Chamras Promptmas |
author_sort |
Jamikorn Suk-Anake |
title |
Analytical performance of centrifuge-based device for clinical chemistry testing |
title_short |
Analytical performance of centrifuge-based device for clinical chemistry testing |
title_full |
Analytical performance of centrifuge-based device for clinical chemistry testing |
title_fullStr |
Analytical performance of centrifuge-based device for clinical chemistry testing |
title_full_unstemmed |
Analytical performance of centrifuge-based device for clinical chemistry testing |
title_sort |
analytical performance of centrifuge-based device for clinical chemistry testing |
publishDate |
2018 |
url |
https://repository.li.mahidol.ac.th/handle/123456789/13542 |
_version_ |
1763493406317740032 |