Optimal monitoring point for temperature and power control in microwave ablation system by using FEM

This study presents an optimal location of monitoring point for using by PID system to control temperature and output power of hepatic cancer microwave ablation system. We investigated a relation between output power and treatment time in hepatic ablation. A simple simulation model was adapted from...

全面介紹

Saved in:
書目詳細資料
Main Authors: S. Pairoch, P. Phasukkit, A. Sanpanich
其他作者: King Mongkut's Institute of Technology Ladkrabang
格式: Conference or Workshop Item
出版: 2018
主題:
在線閱讀:https://repository.li.mahidol.ac.th/handle/123456789/14129
標簽: 添加標簽
沒有標簽, 成為第一個標記此記錄!
機構: Mahidol University
id th-mahidol.14129
record_format dspace
spelling th-mahidol.141292018-06-11T11:47:29Z Optimal monitoring point for temperature and power control in microwave ablation system by using FEM S. Pairoch P. Phasukkit A. Sanpanich King Mongkut's Institute of Technology Ladkrabang Mahidol University Engineering This study presents an optimal location of monitoring point for using by PID system to control temperature and output power of hepatic cancer microwave ablation system. We investigated a relation between output power and treatment time in hepatic ablation. A simple simulation model was adapted from COMSOL FEM solver. As a preliminary, monitoring point was supposed to place at 0.5 centimeter besides ablation probe. The simulation results shown that temperature at this monitoring point is linearly related to output power and treatment time. ©2012 IEEE. 2018-06-11T04:47:29Z 2018-06-11T04:47:29Z 2012-12-01 Conference Paper 5th 2012 Biomedical Engineering International Conference, BMEiCON 2012. (2012) 10.1109/BMEiCon.2012.6465476 2-s2.0-84875094457 https://repository.li.mahidol.ac.th/handle/123456789/14129 Mahidol University SCOPUS https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84875094457&origin=inward
institution Mahidol University
building Mahidol University Library
continent Asia
country Thailand
Thailand
content_provider Mahidol University Library
collection Mahidol University Institutional Repository
topic Engineering
spellingShingle Engineering
S. Pairoch
P. Phasukkit
A. Sanpanich
Optimal monitoring point for temperature and power control in microwave ablation system by using FEM
description This study presents an optimal location of monitoring point for using by PID system to control temperature and output power of hepatic cancer microwave ablation system. We investigated a relation between output power and treatment time in hepatic ablation. A simple simulation model was adapted from COMSOL FEM solver. As a preliminary, monitoring point was supposed to place at 0.5 centimeter besides ablation probe. The simulation results shown that temperature at this monitoring point is linearly related to output power and treatment time. ©2012 IEEE.
author2 King Mongkut's Institute of Technology Ladkrabang
author_facet King Mongkut's Institute of Technology Ladkrabang
S. Pairoch
P. Phasukkit
A. Sanpanich
format Conference or Workshop Item
author S. Pairoch
P. Phasukkit
A. Sanpanich
author_sort S. Pairoch
title Optimal monitoring point for temperature and power control in microwave ablation system by using FEM
title_short Optimal monitoring point for temperature and power control in microwave ablation system by using FEM
title_full Optimal monitoring point for temperature and power control in microwave ablation system by using FEM
title_fullStr Optimal monitoring point for temperature and power control in microwave ablation system by using FEM
title_full_unstemmed Optimal monitoring point for temperature and power control in microwave ablation system by using FEM
title_sort optimal monitoring point for temperature and power control in microwave ablation system by using fem
publishDate 2018
url https://repository.li.mahidol.ac.th/handle/123456789/14129
_version_ 1763492609723990016