Evaluation of a random access analyser: BM/Hitachi 911

The performance of Boehringer Mannheim’s BM/Hitachi 911 was evaluated for three months. The mean coefficient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1-16% (range 0’47-2’38%) and 1-36% (range 0-62-2-93%), respectively. A linearity study for th...

Full description

Saved in:
Bibliographic Details
Main Authors: Taweesook Kanluan, Surapon Tangvarasittichai, Orathai Tangvarasittichai
Other Authors: Mahidol University
Format: Article
Published: 2018
Subjects:
Online Access:https://repository.li.mahidol.ac.th/handle/123456789/17254
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Mahidol University
id th-mahidol.17254
record_format dspace
spelling th-mahidol.172542018-07-04T13:52:21Z Evaluation of a random access analyser: BM/Hitachi 911 Taweesook Kanluan Surapon Tangvarasittichai Orathai Tangvarasittichai Mahidol University Biochemistry, Genetics and Molecular Biology The performance of Boehringer Mannheim’s BM/Hitachi 911 was evaluated for three months. The mean coefficient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1-16% (range 0’47-2’38%) and 1-36% (range 0-62-2-93%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an eight-hour assay nor was any sample-related carry-over detected. In all cases, the regression analyses (slopes) of the results obtained from BM/Hitachi 911 and 717 were between the extreme values of 0-94 and 1-05. During the three months of operation, no major problem was encountered. The BM/Hitachi 911 was found to be easily operated, to require minimal attention and simple daily maintenance during operation. © 1995, Hindawi Publishing Corporation. All rights reserved. 2018-07-04T06:52:21Z 2018-07-04T06:52:21Z 1995-01-01 Article Journal of Automatic Chemistry. Vol.17, No.6 (1995), 213-218 10.1155/S1463924695000332 01420453 2-s2.0-0029590117 https://repository.li.mahidol.ac.th/handle/123456789/17254 Mahidol University SCOPUS https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=0029590117&origin=inward
institution Mahidol University
building Mahidol University Library
continent Asia
country Thailand
Thailand
content_provider Mahidol University Library
collection Mahidol University Institutional Repository
topic Biochemistry, Genetics and Molecular Biology
spellingShingle Biochemistry, Genetics and Molecular Biology
Taweesook Kanluan
Surapon Tangvarasittichai
Orathai Tangvarasittichai
Evaluation of a random access analyser: BM/Hitachi 911
description The performance of Boehringer Mannheim’s BM/Hitachi 911 was evaluated for three months. The mean coefficient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1-16% (range 0’47-2’38%) and 1-36% (range 0-62-2-93%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an eight-hour assay nor was any sample-related carry-over detected. In all cases, the regression analyses (slopes) of the results obtained from BM/Hitachi 911 and 717 were between the extreme values of 0-94 and 1-05. During the three months of operation, no major problem was encountered. The BM/Hitachi 911 was found to be easily operated, to require minimal attention and simple daily maintenance during operation. © 1995, Hindawi Publishing Corporation. All rights reserved.
author2 Mahidol University
author_facet Mahidol University
Taweesook Kanluan
Surapon Tangvarasittichai
Orathai Tangvarasittichai
format Article
author Taweesook Kanluan
Surapon Tangvarasittichai
Orathai Tangvarasittichai
author_sort Taweesook Kanluan
title Evaluation of a random access analyser: BM/Hitachi 911
title_short Evaluation of a random access analyser: BM/Hitachi 911
title_full Evaluation of a random access analyser: BM/Hitachi 911
title_fullStr Evaluation of a random access analyser: BM/Hitachi 911
title_full_unstemmed Evaluation of a random access analyser: BM/Hitachi 911
title_sort evaluation of a random access analyser: bm/hitachi 911
publishDate 2018
url https://repository.li.mahidol.ac.th/handle/123456789/17254
_version_ 1763490499975446528