Evaluation of a random access analyser: BM/Hitachi 911
The performance of Boehringer Mannheim’s BM/Hitachi 911 was evaluated for three months. The mean coefficient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1-16% (range 0’47-2’38%) and 1-36% (range 0-62-2-93%), respectively. A linearity study for th...
Saved in:
Main Authors: | , , |
---|---|
Other Authors: | |
Format: | Article |
Published: |
2018
|
Subjects: | |
Online Access: | https://repository.li.mahidol.ac.th/handle/123456789/17254 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Mahidol University |
id |
th-mahidol.17254 |
---|---|
record_format |
dspace |
spelling |
th-mahidol.172542018-07-04T13:52:21Z Evaluation of a random access analyser: BM/Hitachi 911 Taweesook Kanluan Surapon Tangvarasittichai Orathai Tangvarasittichai Mahidol University Biochemistry, Genetics and Molecular Biology The performance of Boehringer Mannheim’s BM/Hitachi 911 was evaluated for three months. The mean coefficient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1-16% (range 0’47-2’38%) and 1-36% (range 0-62-2-93%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an eight-hour assay nor was any sample-related carry-over detected. In all cases, the regression analyses (slopes) of the results obtained from BM/Hitachi 911 and 717 were between the extreme values of 0-94 and 1-05. During the three months of operation, no major problem was encountered. The BM/Hitachi 911 was found to be easily operated, to require minimal attention and simple daily maintenance during operation. © 1995, Hindawi Publishing Corporation. All rights reserved. 2018-07-04T06:52:21Z 2018-07-04T06:52:21Z 1995-01-01 Article Journal of Automatic Chemistry. Vol.17, No.6 (1995), 213-218 10.1155/S1463924695000332 01420453 2-s2.0-0029590117 https://repository.li.mahidol.ac.th/handle/123456789/17254 Mahidol University SCOPUS https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=0029590117&origin=inward |
institution |
Mahidol University |
building |
Mahidol University Library |
continent |
Asia |
country |
Thailand Thailand |
content_provider |
Mahidol University Library |
collection |
Mahidol University Institutional Repository |
topic |
Biochemistry, Genetics and Molecular Biology |
spellingShingle |
Biochemistry, Genetics and Molecular Biology Taweesook Kanluan Surapon Tangvarasittichai Orathai Tangvarasittichai Evaluation of a random access analyser: BM/Hitachi 911 |
description |
The performance of Boehringer Mannheim’s BM/Hitachi 911 was evaluated for three months. The mean coefficient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1-16% (range 0’47-2’38%) and 1-36% (range 0-62-2-93%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an eight-hour assay nor was any sample-related carry-over detected. In all cases, the regression analyses (slopes) of the results obtained from BM/Hitachi 911 and 717 were between the extreme values of 0-94 and 1-05. During the three months of operation, no major problem was encountered. The BM/Hitachi 911 was found to be easily operated, to require minimal attention and simple daily maintenance during operation. © 1995, Hindawi Publishing Corporation. All rights reserved. |
author2 |
Mahidol University |
author_facet |
Mahidol University Taweesook Kanluan Surapon Tangvarasittichai Orathai Tangvarasittichai |
format |
Article |
author |
Taweesook Kanluan Surapon Tangvarasittichai Orathai Tangvarasittichai |
author_sort |
Taweesook Kanluan |
title |
Evaluation of a random access analyser: BM/Hitachi 911 |
title_short |
Evaluation of a random access analyser: BM/Hitachi 911 |
title_full |
Evaluation of a random access analyser: BM/Hitachi 911 |
title_fullStr |
Evaluation of a random access analyser: BM/Hitachi 911 |
title_full_unstemmed |
Evaluation of a random access analyser: BM/Hitachi 911 |
title_sort |
evaluation of a random access analyser: bm/hitachi 911 |
publishDate |
2018 |
url |
https://repository.li.mahidol.ac.th/handle/123456789/17254 |
_version_ |
1763490499975446528 |