Method performance study of on-line internal standard addition in inductively coupled plasma mass spectrometry

The approach of adding an internal standard to all solutions at a constant level, by on-line merging of the internal standard (steady state) and the sample solution streams, was studied for multi element determinations using inductively coupled plasma mass spectrometry. An internal standard introduc...

Full description

Saved in:
Bibliographic Details
Main Authors: András Törvényi, Kunchit Judprasong, Aleš Fajgelj
Other Authors: International Atomic Energy Agency, Vienna
Format: Article
Published: 2018
Subjects:
Online Access:https://repository.li.mahidol.ac.th/handle/123456789/24340
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Mahidol University
id th-mahidol.24340
record_format dspace
spelling th-mahidol.243402018-08-24T08:46:08Z Method performance study of on-line internal standard addition in inductively coupled plasma mass spectrometry András Törvényi Kunchit Judprasong Aleš Fajgelj International Atomic Energy Agency, Vienna Mahidol University Chemistry The approach of adding an internal standard to all solutions at a constant level, by on-line merging of the internal standard (steady state) and the sample solution streams, was studied for multi element determinations using inductively coupled plasma mass spectrometry. An internal standard introduction system was developed using a knotted reactor, a mixing T-piece and two peristaltic pump channels. The tube length of the knotted reactor and the wash out time were optimized for achieving the best mixing efficiency. The performance of the optimized system was compared to the traditional method of internal standard addition (batch) and checked by analysis of the NIST 8415 SRM (egg powder). The uncertainty components of the internal standard addition were also evaluated and its contribution to the combined measurement uncertainty estimated. No significant difference was found in terms of accuracy and precision between the two methods. However, the new technique was proved to be very practical, robust and less time consuming than the traditional approach. © The Royal Society of Chemistry 2007. 2018-08-24T01:46:08Z 2018-08-24T01:46:08Z 2007-08-30 Article Journal of Analytical Atomic Spectrometry. Vol.22, No.9 (2007), 1152-1157 10.1039/b704867j 13645544 02679477 2-s2.0-34548150403 https://repository.li.mahidol.ac.th/handle/123456789/24340 Mahidol University SCOPUS https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=34548150403&origin=inward
institution Mahidol University
building Mahidol University Library
continent Asia
country Thailand
Thailand
content_provider Mahidol University Library
collection Mahidol University Institutional Repository
topic Chemistry
spellingShingle Chemistry
András Törvényi
Kunchit Judprasong
Aleš Fajgelj
Method performance study of on-line internal standard addition in inductively coupled plasma mass spectrometry
description The approach of adding an internal standard to all solutions at a constant level, by on-line merging of the internal standard (steady state) and the sample solution streams, was studied for multi element determinations using inductively coupled plasma mass spectrometry. An internal standard introduction system was developed using a knotted reactor, a mixing T-piece and two peristaltic pump channels. The tube length of the knotted reactor and the wash out time were optimized for achieving the best mixing efficiency. The performance of the optimized system was compared to the traditional method of internal standard addition (batch) and checked by analysis of the NIST 8415 SRM (egg powder). The uncertainty components of the internal standard addition were also evaluated and its contribution to the combined measurement uncertainty estimated. No significant difference was found in terms of accuracy and precision between the two methods. However, the new technique was proved to be very practical, robust and less time consuming than the traditional approach. © The Royal Society of Chemistry 2007.
author2 International Atomic Energy Agency, Vienna
author_facet International Atomic Energy Agency, Vienna
András Törvényi
Kunchit Judprasong
Aleš Fajgelj
format Article
author András Törvényi
Kunchit Judprasong
Aleš Fajgelj
author_sort András Törvényi
title Method performance study of on-line internal standard addition in inductively coupled plasma mass spectrometry
title_short Method performance study of on-line internal standard addition in inductively coupled plasma mass spectrometry
title_full Method performance study of on-line internal standard addition in inductively coupled plasma mass spectrometry
title_fullStr Method performance study of on-line internal standard addition in inductively coupled plasma mass spectrometry
title_full_unstemmed Method performance study of on-line internal standard addition in inductively coupled plasma mass spectrometry
title_sort method performance study of on-line internal standard addition in inductively coupled plasma mass spectrometry
publishDate 2018
url https://repository.li.mahidol.ac.th/handle/123456789/24340
_version_ 1763488216698060800