SU‐GG‐T‐615: Evaluation of Missing Tissue Compensators for Conventional Dose Calculation Techniques on Pinnacle Treatment Planning System Version 7.6c

Purpose: To validate the accuracy of missing tissue compensators in conventional techniques calculated by the Pinnacle 3D treatment planning system version 7.6c. Method and Materials: For the Pinnacle version 7.6c, the density in g/cm3of lead material and the distance from the source to the top of t...

Full description

Saved in:
Bibliographic Details
Main Authors: P. Changkaew, P. Tangboonduangjit, S. Khachonkom
Other Authors: Mahidol University
Format: Conference or Workshop Item
Published: 2018
Subjects:
Online Access:https://repository.li.mahidol.ac.th/handle/123456789/28818
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Mahidol University
id th-mahidol.28818
record_format dspace
spelling th-mahidol.288182018-09-24T16:36:48Z SU‐GG‐T‐615: Evaluation of Missing Tissue Compensators for Conventional Dose Calculation Techniques on Pinnacle Treatment Planning System Version 7.6c P. Changkaew P. Tangboonduangjit S. Khachonkom Mahidol University Biochemistry, Genetics and Molecular Biology Medicine Purpose: To validate the accuracy of missing tissue compensators in conventional techniques calculated by the Pinnacle 3D treatment planning system version 7.6c. Method and Materials: For the Pinnacle version 7.6c, the density in g/cm3of lead material and the distance from the source to the top of the compensator tray were specified. Lead sheet thicknesses were entered manually pixel by pixel to create the compensators dimension and position corresponding to the missing surface until the isodose profiles were uniform. In solid water phantoms, the lead sheet missing tissue compensators were designed in the conventional techniques at depth 5 and 10 cm. The thicknesses of 1, 3, 5 and 7 cm of step solid water phantoms was used to simulate as a missing tissue. Kodak X‐Omat V film was used to verify the ID dose profiles. OmniPro‐I'mRT software analysis was employed for the dose comparison between measurement and calculation. Results: The study showed that, the missing tissue thicknesses of 1, 3, 5 and 7 cm need to be compensated using lead sheets thicknesses of 0.65, 2.06, 3.62 and 5.42 mm, respectively. The difference of all dose profiles between the Pinnacle 3D treatment planning and the film measurement are within 3% which agree in the criteria of TRS 430 protocol. Conclusions: The missing tissue compensators using the Pinnacle 3D treatment planning version 7.6c establish the uniform dose distributions for 1D profile comparison which can be applied in clinical situations. © 2010, American Association of Physicists in Medicine. All rights reserved. 2018-09-24T08:49:05Z 2018-09-24T08:49:05Z 2010-01-01 Conference Paper Medical Physics. Vol.37, No.6 (2010), 3329 10.1118/1.3469016 00942405 2-s2.0-85024786686 https://repository.li.mahidol.ac.th/handle/123456789/28818 Mahidol University SCOPUS https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85024786686&origin=inward
institution Mahidol University
building Mahidol University Library
continent Asia
country Thailand
Thailand
content_provider Mahidol University Library
collection Mahidol University Institutional Repository
topic Biochemistry, Genetics and Molecular Biology
Medicine
spellingShingle Biochemistry, Genetics and Molecular Biology
Medicine
P. Changkaew
P. Tangboonduangjit
S. Khachonkom
SU‐GG‐T‐615: Evaluation of Missing Tissue Compensators for Conventional Dose Calculation Techniques on Pinnacle Treatment Planning System Version 7.6c
description Purpose: To validate the accuracy of missing tissue compensators in conventional techniques calculated by the Pinnacle 3D treatment planning system version 7.6c. Method and Materials: For the Pinnacle version 7.6c, the density in g/cm3of lead material and the distance from the source to the top of the compensator tray were specified. Lead sheet thicknesses were entered manually pixel by pixel to create the compensators dimension and position corresponding to the missing surface until the isodose profiles were uniform. In solid water phantoms, the lead sheet missing tissue compensators were designed in the conventional techniques at depth 5 and 10 cm. The thicknesses of 1, 3, 5 and 7 cm of step solid water phantoms was used to simulate as a missing tissue. Kodak X‐Omat V film was used to verify the ID dose profiles. OmniPro‐I'mRT software analysis was employed for the dose comparison between measurement and calculation. Results: The study showed that, the missing tissue thicknesses of 1, 3, 5 and 7 cm need to be compensated using lead sheets thicknesses of 0.65, 2.06, 3.62 and 5.42 mm, respectively. The difference of all dose profiles between the Pinnacle 3D treatment planning and the film measurement are within 3% which agree in the criteria of TRS 430 protocol. Conclusions: The missing tissue compensators using the Pinnacle 3D treatment planning version 7.6c establish the uniform dose distributions for 1D profile comparison which can be applied in clinical situations. © 2010, American Association of Physicists in Medicine. All rights reserved.
author2 Mahidol University
author_facet Mahidol University
P. Changkaew
P. Tangboonduangjit
S. Khachonkom
format Conference or Workshop Item
author P. Changkaew
P. Tangboonduangjit
S. Khachonkom
author_sort P. Changkaew
title SU‐GG‐T‐615: Evaluation of Missing Tissue Compensators for Conventional Dose Calculation Techniques on Pinnacle Treatment Planning System Version 7.6c
title_short SU‐GG‐T‐615: Evaluation of Missing Tissue Compensators for Conventional Dose Calculation Techniques on Pinnacle Treatment Planning System Version 7.6c
title_full SU‐GG‐T‐615: Evaluation of Missing Tissue Compensators for Conventional Dose Calculation Techniques on Pinnacle Treatment Planning System Version 7.6c
title_fullStr SU‐GG‐T‐615: Evaluation of Missing Tissue Compensators for Conventional Dose Calculation Techniques on Pinnacle Treatment Planning System Version 7.6c
title_full_unstemmed SU‐GG‐T‐615: Evaluation of Missing Tissue Compensators for Conventional Dose Calculation Techniques on Pinnacle Treatment Planning System Version 7.6c
title_sort su‐gg‐t‐615: evaluation of missing tissue compensators for conventional dose calculation techniques on pinnacle treatment planning system version 7.6c
publishDate 2018
url https://repository.li.mahidol.ac.th/handle/123456789/28818
_version_ 1763491968024838144