Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: A cross-sectional method comparison study

© 2020 The Author(s). Background: Transfusion decision during the perioperative period mostly relies on the point-of-care testing for Hb measurement. This study aimed systematically compared four point-of-care methods with the central laboratory measurement of hemoglobin (LHb) regarding the accuracy...

Full description

Saved in:
Bibliographic Details
Main Authors: Arpa Chutipongtanate, Churairat Yasaeng, Tanit Virankabutra, Somchai Chutipongtanate
Other Authors: Faculty of Medicine, Ramathibodi Hospital, Mahidol University
Format: Article
Published: 2020
Subjects:
Online Access:https://repository.li.mahidol.ac.th/handle/123456789/56254
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Mahidol University
id th-mahidol.56254
record_format dspace
spelling th-mahidol.562542020-06-02T11:58:08Z Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: A cross-sectional method comparison study Arpa Chutipongtanate Churairat Yasaeng Tanit Virankabutra Somchai Chutipongtanate Faculty of Medicine, Ramathibodi Hospital, Mahidol University Medicine © 2020 The Author(s). Background: Transfusion decision during the perioperative period mostly relies on the point-of-care testing for Hb measurement. This study aimed systematically compared four point-of-care methods with the central laboratory measurement of hemoglobin (LHb) regarding the accuracy, precision, and assay practicality to identify the preferred point-of-care method during the perioperative period. Methods: This cross-sectional method comparison study was conducted in the surgical intensive care unit at Ramathibodi Hospital, Thailand, from September 2015 to July 2016. Four point-of-care methods, i.e., capillary hematocrit (HctCap), HemoCue Hb201+, iSTAT with CG8+ cartridge, and SpHb from Radical-7 pulse co-oximeter were carried out when LHb was ordered. Pearson correlation and Bland-Altman analyses were performed to assess the accuracy and precision, while the workload, turnaround time, and the unit cost were evaluated for the method practicality. Results: Thirty-five patients were enrolled, corresponding to 48 blood specimens for analyses, resulting in the measured hemoglobin of 11.2 ± 1.9 g/dL by LHb. Ranking by correlation (r), mean difference (bias) and 95% limit of agreement (LOA) showed the point-of-care methods from the greater to the less performance as followed, iSTAT-LHb pair (r = 0.941; bias 0.15 (95% LOA; - 1.41, 1.12) g/dL), HemoCue-LHb pair (r = 0.922; bias - 0.18 (95% LOA; - 1.63, 1.28) g/dL), SpHb-LHb pair (r = 0.670; bias 0.13 (95% LOA; - 3.12, 3.39) g/dL) and HctCap-LHb pair (r = 0.905; bias 0.46 (95% LOA; - 1.16, 2.08) g/dL). Considering the practicality, all point-of-care methods had less workload and turnaround time than LHb, but only HemoCue and HctCap had lower unit cost. Conclusion: This study identified HemoCue as the suitable point-of-care method for the sole purpose of Hb measurement in the surgical ICU setting, while iSTAT should be considered when additional data is needed. 2020-06-02T04:58:08Z 2020-06-02T04:58:08Z 2020-04-22 Article BMC Anesthesiology. Vol.20, No.1 (2020) 10.1186/s12871-020-01008-8 14712253 2-s2.0-85083947399 https://repository.li.mahidol.ac.th/handle/123456789/56254 Mahidol University SCOPUS https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85083947399&origin=inward
institution Mahidol University
building Mahidol University Library
continent Asia
country Thailand
Thailand
content_provider Mahidol University Library
collection Mahidol University Institutional Repository
topic Medicine
spellingShingle Medicine
Arpa Chutipongtanate
Churairat Yasaeng
Tanit Virankabutra
Somchai Chutipongtanate
Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: A cross-sectional method comparison study
description © 2020 The Author(s). Background: Transfusion decision during the perioperative period mostly relies on the point-of-care testing for Hb measurement. This study aimed systematically compared four point-of-care methods with the central laboratory measurement of hemoglobin (LHb) regarding the accuracy, precision, and assay practicality to identify the preferred point-of-care method during the perioperative period. Methods: This cross-sectional method comparison study was conducted in the surgical intensive care unit at Ramathibodi Hospital, Thailand, from September 2015 to July 2016. Four point-of-care methods, i.e., capillary hematocrit (HctCap), HemoCue Hb201+, iSTAT with CG8+ cartridge, and SpHb from Radical-7 pulse co-oximeter were carried out when LHb was ordered. Pearson correlation and Bland-Altman analyses were performed to assess the accuracy and precision, while the workload, turnaround time, and the unit cost were evaluated for the method practicality. Results: Thirty-five patients were enrolled, corresponding to 48 blood specimens for analyses, resulting in the measured hemoglobin of 11.2 ± 1.9 g/dL by LHb. Ranking by correlation (r), mean difference (bias) and 95% limit of agreement (LOA) showed the point-of-care methods from the greater to the less performance as followed, iSTAT-LHb pair (r = 0.941; bias 0.15 (95% LOA; - 1.41, 1.12) g/dL), HemoCue-LHb pair (r = 0.922; bias - 0.18 (95% LOA; - 1.63, 1.28) g/dL), SpHb-LHb pair (r = 0.670; bias 0.13 (95% LOA; - 3.12, 3.39) g/dL) and HctCap-LHb pair (r = 0.905; bias 0.46 (95% LOA; - 1.16, 2.08) g/dL). Considering the practicality, all point-of-care methods had less workload and turnaround time than LHb, but only HemoCue and HctCap had lower unit cost. Conclusion: This study identified HemoCue as the suitable point-of-care method for the sole purpose of Hb measurement in the surgical ICU setting, while iSTAT should be considered when additional data is needed.
author2 Faculty of Medicine, Ramathibodi Hospital, Mahidol University
author_facet Faculty of Medicine, Ramathibodi Hospital, Mahidol University
Arpa Chutipongtanate
Churairat Yasaeng
Tanit Virankabutra
Somchai Chutipongtanate
format Article
author Arpa Chutipongtanate
Churairat Yasaeng
Tanit Virankabutra
Somchai Chutipongtanate
author_sort Arpa Chutipongtanate
title Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: A cross-sectional method comparison study
title_short Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: A cross-sectional method comparison study
title_full Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: A cross-sectional method comparison study
title_fullStr Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: A cross-sectional method comparison study
title_full_unstemmed Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: A cross-sectional method comparison study
title_sort systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical icu setting: a cross-sectional method comparison study
publishDate 2020
url https://repository.li.mahidol.ac.th/handle/123456789/56254
_version_ 1763497045033746432