Performance Evaluation of BD FACSPresto™ Near-Patient CD4 Counter for Monitoring Antiretroviral Therapy in HIV-Infected Individuals in Primary Healthcare Clinics in Thailand

HIV viral load is more reliable tool for monitoring treatment throughout the course of HIV/AIDS, but the test may be expensive in resource-limited settings. Therefore, enumeration of CD4 T-lymphocyte count remains important in these settings. This study evaluated the performance of BDFACSPresto, a n...

Full description

Saved in:
Bibliographic Details
Main Authors: Kasama Sukapirom, Somrat Matchua, Charin Thepthai, Narinee Srimark, Ladawan Khowawisetsut, Kovit Pattanapanyasat
Other Authors: Siriraj Hospital
Format: Article
Published: 2022
Subjects:
Online Access:https://repository.li.mahidol.ac.th/handle/123456789/73443
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Mahidol University
id th-mahidol.73443
record_format dspace
spelling th-mahidol.734432022-08-04T10:43:50Z Performance Evaluation of BD FACSPresto™ Near-Patient CD4 Counter for Monitoring Antiretroviral Therapy in HIV-Infected Individuals in Primary Healthcare Clinics in Thailand Kasama Sukapirom Somrat Matchua Charin Thepthai Narinee Srimark Ladawan Khowawisetsut Kovit Pattanapanyasat Siriraj Hospital Chonburi Regional Hospital Biochemistry, Genetics and Molecular Biology HIV viral load is more reliable tool for monitoring treatment throughout the course of HIV/AIDS, but the test may be expensive in resource-limited settings. Therefore, enumeration of CD4 T-lymphocyte count remains important in these settings. This study evaluated the performance of BDFACSPresto, a near-patient CD4 counter planned to be used in primary healthcare clinics in Thailand. Results of percent, absolute CD4 count and hemoglobin (Hb) on the FACSPresto were compared with the TriTEST/TruCOUNT/BDFACSCalibur method and a Sysmex hematology ana-lyzer. Phase I of the study was performed in an ISO15189 laboratory. Both percentage and absolute values showed Passing–Bablok slopes within 0.98–1.06 and 0.97–1.13, mean Bland–Altman biases of +1.2% and +20.5 cells/µL, respectively. In phase II, venous and some capillary blood samples were analyzed in four primary healthcare clinics. The results showed good correlation between capillary and venous blood. For venous blood samples, regression lines showed slopes of 1.01–1.05 and 1.01– 1.07 for all percentage and absolute values. The overall mean biases were +0.9% and +17.0 cells/µL. For Hb, Passing–Bablok regression result gave slope within 1.01–1.07 and mean bias of −0.06 g/dL. Thus, CD4 enumeration in blood by the FACSPresto is reliable and can be performed to an identical standard at primary healthcare clinics. 2022-08-04T03:43:50Z 2022-08-04T03:43:50Z 2022-02-01 Article Diagnostics. Vol.12, No.2 (2022) 10.3390/diagnostics12020382 20754418 2-s2.0-85124075989 https://repository.li.mahidol.ac.th/handle/123456789/73443 Mahidol University SCOPUS https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85124075989&origin=inward
institution Mahidol University
building Mahidol University Library
continent Asia
country Thailand
Thailand
content_provider Mahidol University Library
collection Mahidol University Institutional Repository
topic Biochemistry, Genetics and Molecular Biology
spellingShingle Biochemistry, Genetics and Molecular Biology
Kasama Sukapirom
Somrat Matchua
Charin Thepthai
Narinee Srimark
Ladawan Khowawisetsut
Kovit Pattanapanyasat
Performance Evaluation of BD FACSPresto™ Near-Patient CD4 Counter for Monitoring Antiretroviral Therapy in HIV-Infected Individuals in Primary Healthcare Clinics in Thailand
description HIV viral load is more reliable tool for monitoring treatment throughout the course of HIV/AIDS, but the test may be expensive in resource-limited settings. Therefore, enumeration of CD4 T-lymphocyte count remains important in these settings. This study evaluated the performance of BDFACSPresto, a near-patient CD4 counter planned to be used in primary healthcare clinics in Thailand. Results of percent, absolute CD4 count and hemoglobin (Hb) on the FACSPresto were compared with the TriTEST/TruCOUNT/BDFACSCalibur method and a Sysmex hematology ana-lyzer. Phase I of the study was performed in an ISO15189 laboratory. Both percentage and absolute values showed Passing–Bablok slopes within 0.98–1.06 and 0.97–1.13, mean Bland–Altman biases of +1.2% and +20.5 cells/µL, respectively. In phase II, venous and some capillary blood samples were analyzed in four primary healthcare clinics. The results showed good correlation between capillary and venous blood. For venous blood samples, regression lines showed slopes of 1.01–1.05 and 1.01– 1.07 for all percentage and absolute values. The overall mean biases were +0.9% and +17.0 cells/µL. For Hb, Passing–Bablok regression result gave slope within 1.01–1.07 and mean bias of −0.06 g/dL. Thus, CD4 enumeration in blood by the FACSPresto is reliable and can be performed to an identical standard at primary healthcare clinics.
author2 Siriraj Hospital
author_facet Siriraj Hospital
Kasama Sukapirom
Somrat Matchua
Charin Thepthai
Narinee Srimark
Ladawan Khowawisetsut
Kovit Pattanapanyasat
format Article
author Kasama Sukapirom
Somrat Matchua
Charin Thepthai
Narinee Srimark
Ladawan Khowawisetsut
Kovit Pattanapanyasat
author_sort Kasama Sukapirom
title Performance Evaluation of BD FACSPresto™ Near-Patient CD4 Counter for Monitoring Antiretroviral Therapy in HIV-Infected Individuals in Primary Healthcare Clinics in Thailand
title_short Performance Evaluation of BD FACSPresto™ Near-Patient CD4 Counter for Monitoring Antiretroviral Therapy in HIV-Infected Individuals in Primary Healthcare Clinics in Thailand
title_full Performance Evaluation of BD FACSPresto™ Near-Patient CD4 Counter for Monitoring Antiretroviral Therapy in HIV-Infected Individuals in Primary Healthcare Clinics in Thailand
title_fullStr Performance Evaluation of BD FACSPresto™ Near-Patient CD4 Counter for Monitoring Antiretroviral Therapy in HIV-Infected Individuals in Primary Healthcare Clinics in Thailand
title_full_unstemmed Performance Evaluation of BD FACSPresto™ Near-Patient CD4 Counter for Monitoring Antiretroviral Therapy in HIV-Infected Individuals in Primary Healthcare Clinics in Thailand
title_sort performance evaluation of bd facspresto™ near-patient cd4 counter for monitoring antiretroviral therapy in hiv-infected individuals in primary healthcare clinics in thailand
publishDate 2022
url https://repository.li.mahidol.ac.th/handle/123456789/73443
_version_ 1763489859163389952