Quantitative Determination of Iron-Siderophore Complexes in Peat by Isotope-Exchange Size-Exclusion UPLC-Electrospray Ionization High-Resolution Accurate Mass (HRAM) Mass Spectrometry

A method was developed for the quantification of iron-siderophore complexes by electrospray high-resolution accurate mass (HRAM) mass spectrometry (MS) without the need for authentic standards. The bulk of iron-siderophore complexes was purified by solid-phase extraction (SPE) and concentrated by ev...

Full description

Saved in:
Bibliographic Details
Main Author: Maknun L.
Other Authors: Mahidol University
Format: Article
Published: 2023
Subjects:
Online Access:https://repository.li.mahidol.ac.th/handle/123456789/87987
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Mahidol University
id th-mahidol.87987
record_format dspace
spelling th-mahidol.879872023-07-21T01:01:16Z Quantitative Determination of Iron-Siderophore Complexes in Peat by Isotope-Exchange Size-Exclusion UPLC-Electrospray Ionization High-Resolution Accurate Mass (HRAM) Mass Spectrometry Maknun L. Mahidol University Chemistry A method was developed for the quantification of iron-siderophore complexes by electrospray high-resolution accurate mass (HRAM) mass spectrometry (MS) without the need for authentic standards. The bulk of iron-siderophore complexes was purified by solid-phase extraction (SPE) and concentrated by evaporation. The individual complexes were identified by fast size-exclusion chromatography (FastSEC)-Orbitrap MSn on the basis of the exact molecular mass (±1 ppm) and MS2 or MS3 fragmentation. Their capability to readily exchange the natural 56Fe for the added 58Fe was demonstrated by SEC with ICP MS and ESI MS detection. The method was applied to the analysis of peat sampled in the eastern part of the French Pyrenean mountains. Nineteen siderophores belonging to four different classes were identified and quantified. The results were validated using ICP MS detection of iron by matching the sum of iron complexes determined by isotope exchange-ESI MS within each peak observed by FastSEC-ICP MS. 2023-07-20T18:01:16Z 2023-07-20T18:01:16Z 2023-01-01 Article Analytical Chemistry (2023) 10.1021/acs.analchem.3c00122 15206882 00032700 2-s2.0-85163720321 https://repository.li.mahidol.ac.th/handle/123456789/87987 SCOPUS
institution Mahidol University
building Mahidol University Library
continent Asia
country Thailand
Thailand
content_provider Mahidol University Library
collection Mahidol University Institutional Repository
topic Chemistry
spellingShingle Chemistry
Maknun L.
Quantitative Determination of Iron-Siderophore Complexes in Peat by Isotope-Exchange Size-Exclusion UPLC-Electrospray Ionization High-Resolution Accurate Mass (HRAM) Mass Spectrometry
description A method was developed for the quantification of iron-siderophore complexes by electrospray high-resolution accurate mass (HRAM) mass spectrometry (MS) without the need for authentic standards. The bulk of iron-siderophore complexes was purified by solid-phase extraction (SPE) and concentrated by evaporation. The individual complexes were identified by fast size-exclusion chromatography (FastSEC)-Orbitrap MSn on the basis of the exact molecular mass (±1 ppm) and MS2 or MS3 fragmentation. Their capability to readily exchange the natural 56Fe for the added 58Fe was demonstrated by SEC with ICP MS and ESI MS detection. The method was applied to the analysis of peat sampled in the eastern part of the French Pyrenean mountains. Nineteen siderophores belonging to four different classes were identified and quantified. The results were validated using ICP MS detection of iron by matching the sum of iron complexes determined by isotope exchange-ESI MS within each peak observed by FastSEC-ICP MS.
author2 Mahidol University
author_facet Mahidol University
Maknun L.
format Article
author Maknun L.
author_sort Maknun L.
title Quantitative Determination of Iron-Siderophore Complexes in Peat by Isotope-Exchange Size-Exclusion UPLC-Electrospray Ionization High-Resolution Accurate Mass (HRAM) Mass Spectrometry
title_short Quantitative Determination of Iron-Siderophore Complexes in Peat by Isotope-Exchange Size-Exclusion UPLC-Electrospray Ionization High-Resolution Accurate Mass (HRAM) Mass Spectrometry
title_full Quantitative Determination of Iron-Siderophore Complexes in Peat by Isotope-Exchange Size-Exclusion UPLC-Electrospray Ionization High-Resolution Accurate Mass (HRAM) Mass Spectrometry
title_fullStr Quantitative Determination of Iron-Siderophore Complexes in Peat by Isotope-Exchange Size-Exclusion UPLC-Electrospray Ionization High-Resolution Accurate Mass (HRAM) Mass Spectrometry
title_full_unstemmed Quantitative Determination of Iron-Siderophore Complexes in Peat by Isotope-Exchange Size-Exclusion UPLC-Electrospray Ionization High-Resolution Accurate Mass (HRAM) Mass Spectrometry
title_sort quantitative determination of iron-siderophore complexes in peat by isotope-exchange size-exclusion uplc-electrospray ionization high-resolution accurate mass (hram) mass spectrometry
publishDate 2023
url https://repository.li.mahidol.ac.th/handle/123456789/87987
_version_ 1781415212309544960