Closed-Loop ventilation using sidestream versus mainstream capnography for automated adjustments of minute ventilation-A randomized clinical trial in cardiac surgery patients

BACKGROUND: INTELLiVENT-Adaptive Support Ventilation (ASV) is a closed-loop ventilation mode that uses capnography to adjust tidal volume (VT) and respiratory rate according to a user-set end-tidal CO2 (etCO2) target range. We compared sidestream versus mainstream capnography with this ventilation m...

Full description

Saved in:
Bibliographic Details
Main Author: Nijbroek S.G.L.H.
Other Authors: Mahidol University
Format: Article
Published: 2023
Subjects:
Online Access:https://repository.li.mahidol.ac.th/handle/123456789/89149
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Mahidol University
id th-mahidol.89149
record_format dspace
spelling th-mahidol.891492023-09-03T01:02:18Z Closed-Loop ventilation using sidestream versus mainstream capnography for automated adjustments of minute ventilation-A randomized clinical trial in cardiac surgery patients Nijbroek S.G.L.H. Mahidol University Multidisciplinary BACKGROUND: INTELLiVENT-Adaptive Support Ventilation (ASV) is a closed-loop ventilation mode that uses capnography to adjust tidal volume (VT) and respiratory rate according to a user-set end-tidal CO2 (etCO2) target range. We compared sidestream versus mainstream capnography with this ventilation mode with respect to the quality of breathing in patients after cardiac surgery. METHODS: Single-center, single-blinded, non-inferiority, randomized clinical trial in adult patients scheduled for elective cardiac surgery that were expected to receive at least two hours of postoperative ventilation in the ICU. Patients were randomized 1:1 to closed-loop ventilation with sidestream or mainstream capnography. Each breath was classified into a zone based on the measured VT, maximum airway pressure, etCO2 and pulse oximetry. The primary outcome was the proportion of breaths spent in a predefined 'optimal' zone of ventilation during the first three hours of postoperative ventilation, with a non-inferiority margin for the difference in the proportions set at -20%. Secondary endpoints included the proportion of breaths in predefined 'acceptable' and 'critical' zones of ventilation, and the proportion of breaths with hypoxemia. RESULTS: Of 80 randomized subjects, 78 were included in the intention-to-treat analysis. We could not confirm the non-inferiority of closed-loop ventilation using sidestream with respect to the proportion of breaths in the 'optimal' zone (mean ratio 0.87 [0.77 to ∞]; P = 0.116 for non-inferiority). The proportion of breaths with hypoxemia was higher in the sidestream capnography group versus the mainstream capnography group. CONCLUSIONS: We could not confirm that INTELLiVENT-ASV using sidestream capnography is non-inferior to INTELLiVENT-ASV using mainstream capnography with respect to the quality of breathing in subjects receiving postoperative ventilation after cardiac surgery. TRIAL REGISTRATION: NCT04599491 (clinicaltrials.gov). 2023-09-02T18:02:18Z 2023-09-02T18:02:18Z 2023-01-01 Article PloS one Vol.18 No.8 (2023) , e0289412 10.1371/journal.pone.0289412 19326203 37611007 2-s2.0-85168564413 https://repository.li.mahidol.ac.th/handle/123456789/89149 SCOPUS
institution Mahidol University
building Mahidol University Library
continent Asia
country Thailand
Thailand
content_provider Mahidol University Library
collection Mahidol University Institutional Repository
topic Multidisciplinary
spellingShingle Multidisciplinary
Nijbroek S.G.L.H.
Closed-Loop ventilation using sidestream versus mainstream capnography for automated adjustments of minute ventilation-A randomized clinical trial in cardiac surgery patients
description BACKGROUND: INTELLiVENT-Adaptive Support Ventilation (ASV) is a closed-loop ventilation mode that uses capnography to adjust tidal volume (VT) and respiratory rate according to a user-set end-tidal CO2 (etCO2) target range. We compared sidestream versus mainstream capnography with this ventilation mode with respect to the quality of breathing in patients after cardiac surgery. METHODS: Single-center, single-blinded, non-inferiority, randomized clinical trial in adult patients scheduled for elective cardiac surgery that were expected to receive at least two hours of postoperative ventilation in the ICU. Patients were randomized 1:1 to closed-loop ventilation with sidestream or mainstream capnography. Each breath was classified into a zone based on the measured VT, maximum airway pressure, etCO2 and pulse oximetry. The primary outcome was the proportion of breaths spent in a predefined 'optimal' zone of ventilation during the first three hours of postoperative ventilation, with a non-inferiority margin for the difference in the proportions set at -20%. Secondary endpoints included the proportion of breaths in predefined 'acceptable' and 'critical' zones of ventilation, and the proportion of breaths with hypoxemia. RESULTS: Of 80 randomized subjects, 78 were included in the intention-to-treat analysis. We could not confirm the non-inferiority of closed-loop ventilation using sidestream with respect to the proportion of breaths in the 'optimal' zone (mean ratio 0.87 [0.77 to ∞]; P = 0.116 for non-inferiority). The proportion of breaths with hypoxemia was higher in the sidestream capnography group versus the mainstream capnography group. CONCLUSIONS: We could not confirm that INTELLiVENT-ASV using sidestream capnography is non-inferior to INTELLiVENT-ASV using mainstream capnography with respect to the quality of breathing in subjects receiving postoperative ventilation after cardiac surgery. TRIAL REGISTRATION: NCT04599491 (clinicaltrials.gov).
author2 Mahidol University
author_facet Mahidol University
Nijbroek S.G.L.H.
format Article
author Nijbroek S.G.L.H.
author_sort Nijbroek S.G.L.H.
title Closed-Loop ventilation using sidestream versus mainstream capnography for automated adjustments of minute ventilation-A randomized clinical trial in cardiac surgery patients
title_short Closed-Loop ventilation using sidestream versus mainstream capnography for automated adjustments of minute ventilation-A randomized clinical trial in cardiac surgery patients
title_full Closed-Loop ventilation using sidestream versus mainstream capnography for automated adjustments of minute ventilation-A randomized clinical trial in cardiac surgery patients
title_fullStr Closed-Loop ventilation using sidestream versus mainstream capnography for automated adjustments of minute ventilation-A randomized clinical trial in cardiac surgery patients
title_full_unstemmed Closed-Loop ventilation using sidestream versus mainstream capnography for automated adjustments of minute ventilation-A randomized clinical trial in cardiac surgery patients
title_sort closed-loop ventilation using sidestream versus mainstream capnography for automated adjustments of minute ventilation-a randomized clinical trial in cardiac surgery patients
publishDate 2023
url https://repository.li.mahidol.ac.th/handle/123456789/89149
_version_ 1781416105151037440