#TITLE_ALTERNATIVE#

Ferroelectric hysteresis measurement system has been built to investigate hysteresis curve on ferroelectric materials. In this system, Sawyer-Tower circuit, consisting of linear capacitor and ferroelectric capacitor is used for detecting polarization on ferroelectric materials, and microcontroller C...

Full description

Saved in:
Bibliographic Details
Main Author: WAHYUDHIN FATHONA (NIM 10205023), INDRA
Format: Final Project
Language:Indonesia
Online Access:https://digilib.itb.ac.id/gdl/view/10667
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Institut Teknologi Bandung
Language: Indonesia
Description
Summary:Ferroelectric hysteresis measurement system has been built to investigate hysteresis curve on ferroelectric materials. In this system, Sawyer-Tower circuit, consisting of linear capacitor and ferroelectric capacitor is used for detecting polarization on ferroelectric materials, and microcontroller C8051F005 as main brain controlling the work of the system . Measurement of PZT-ferroelectric materials had been performed by Sawyer-Tower circuit and the result compared with measurement result performed by Radian Rt66A. The aim of this comparison is to observe the accuracy of the system in performed the measurement. The hysteresis curve displayed on Oscilloscope in X-Y mode and Personal Computer with Delphi as Graphic User Interface. Several PZT samples had been measured by this system and the result compared with Radian measurement. <br />