PREPARATION AND CHARACTERIZATION COATED WIRE ISE CHLORIDE

Coated wire chloride ion-selective electrode has been constructed using Aliquat 336-chloride immobilized in polyvinylchloride (PVC) as supporting membrane, and dioctylphtalate (DOP) as a plasticizer. The optimum composition of the membrane materials is 7% of aliquat 336-chloride, 58% of DOP...

Full description

Saved in:
Bibliographic Details
Main Author: Dharmanto, Agung
Format: Final Project
Language:Indonesia
Subjects:
Online Access:https://digilib.itb.ac.id/gdl/view/35140
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Institut Teknologi Bandung
Language: Indonesia
id id-itb.:35140
spelling id-itb.:351402019-02-21T08:38:49ZPREPARATION AND CHARACTERIZATION COATED WIRE ISE CHLORIDE Dharmanto, Agung Kimia Indonesia Final Project ion selective electrodes, coated wire, aliquat 336-chloride, polyvinilchloride, sensors INSTITUT TEKNOLOGI BANDUNG https://digilib.itb.ac.id/gdl/view/35140 Coated wire chloride ion-selective electrode has been constructed using Aliquat 336-chloride immobilized in polyvinylchloride (PVC) as supporting membrane, and dioctylphtalate (DOP) as a plasticizer. The optimum composition of the membrane materials is 7% of aliquat 336-chloride, 58% of DOP and 35% of PVC. The dynamic range of the electrode is between 10-2 –10-4 M of chloride ion, with the best Nernst factor electrode 55,45 mV/decade. Within the dynamic range, the response time of the electrode is between 300 seconds. The optimum pH for the measurement is 4,1. The performance of coated wire electrodes is slightly affected by monovalen ion nitrate. text
institution Institut Teknologi Bandung
building Institut Teknologi Bandung Library
continent Asia
country Indonesia
Indonesia
content_provider Institut Teknologi Bandung
collection Digital ITB
language Indonesia
topic Kimia
spellingShingle Kimia
Dharmanto, Agung
PREPARATION AND CHARACTERIZATION COATED WIRE ISE CHLORIDE
description Coated wire chloride ion-selective electrode has been constructed using Aliquat 336-chloride immobilized in polyvinylchloride (PVC) as supporting membrane, and dioctylphtalate (DOP) as a plasticizer. The optimum composition of the membrane materials is 7% of aliquat 336-chloride, 58% of DOP and 35% of PVC. The dynamic range of the electrode is between 10-2 –10-4 M of chloride ion, with the best Nernst factor electrode 55,45 mV/decade. Within the dynamic range, the response time of the electrode is between 300 seconds. The optimum pH for the measurement is 4,1. The performance of coated wire electrodes is slightly affected by monovalen ion nitrate.
format Final Project
author Dharmanto, Agung
author_facet Dharmanto, Agung
author_sort Dharmanto, Agung
title PREPARATION AND CHARACTERIZATION COATED WIRE ISE CHLORIDE
title_short PREPARATION AND CHARACTERIZATION COATED WIRE ISE CHLORIDE
title_full PREPARATION AND CHARACTERIZATION COATED WIRE ISE CHLORIDE
title_fullStr PREPARATION AND CHARACTERIZATION COATED WIRE ISE CHLORIDE
title_full_unstemmed PREPARATION AND CHARACTERIZATION COATED WIRE ISE CHLORIDE
title_sort preparation and characterization coated wire ise chloride
url https://digilib.itb.ac.id/gdl/view/35140
_version_ 1821996875064016896