UJI VALIDITAS DAN RELIABILITAS BEBERAPA ALAT SWA-MONITOR KOMERSIAL UNTUK PEMERIKSAAN KADAR GLUKOSA DARAH DAN STUDI PENGARUH FAKTOR LOKASI PENGAMBILAN SAMPEL TERHADAP HASIL PENGUKURAN
Blood glucose monitoring is the most routine general checkup that been done independently by society regarding the incidence and prevalence of diabetes mellitus, particularly type 2. Recently, there are many self-monitoring devices for blood glucose, known as glucometer which globally market. Thi...
Saved in:
Main Author: | |
---|---|
Format: | Final Project |
Language: | Indonesia |
Online Access: | https://digilib.itb.ac.id/gdl/view/40294 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Institut Teknologi Bandung |
Language: | Indonesia |
Summary: | Blood glucose monitoring is the most routine general checkup that been done independently by
society regarding the incidence and prevalence of diabetes mellitus, particularly type 2. Recently,
there are many self-monitoring devices for blood glucose, known as glucometer which globally
market. This study aimed to determine the validity of glucometers from several brands compared to
the standard method spectrophotometer which is done by clinical laboratory, determine the reliability
of the glucometers, examine the impact of the location of sample collection to the result of the blood
glucose measurement and also developed the information and appropriate recommendation of
glucometer usage. This study was done to five subjects to compare the measurement of blood glucose
using the glucometers and spectrophotometer as a standard method. Within the same time, sample
collection is done on vein and capillary blood. On this study, blood samples also been taken from five
subjects for twelve times in a month on the right and left hand to examine the reliability and to see
any effect from sampling location. The result is statistically analyzed using the paired t-test and
Cronbach's alpha. The result showed that there are no significant differences between the
measurement of blood glucose using glucometers and spectrophotometer with the p-value in brand
ET, AC, and NES are 0.0934, 0.0766, 0.6926 (>0.05) respectively. Next, the result of blood glucose
measurement between left and right hand is significantly different with P-value in brand ET, AC, and
NES are P<0.0001 (<0.05) and established similar repeatability of results on both hands. On reliability
test, the Chronbach’s alpha value for glucometer brand ET, AC, and NES on daytime measurement are
0.6964, 0.7208, and 0.5459 respectively. Meanwhile, on night-time, the alpha value is resulted in the
yield 0.7814, 0.7426, and 0.7703. This study leading to the conclusion that the glucometers are valid
compared to the standard method and the reliability of the glucometer is better at night-time
measurement while the result of right hand blood glucose is higher than left hand meaning that the
location of sample collection affects the results of the measurement.
|
---|