Analytical evaluation of cardiac poct: humasis hubi-quanpro troponin I
Point-of-care testing (POCT) of cardiac troponin device is aimed for improvement in turn round time (TAT) and assist in acute management care of acute coronary syndrome (ACS). The present study was conducted to assess the analytical performance and correlation of HUBI-QUANPro troponin I with an ex...
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Pusat Perubatan Universiti Kebangsaan Malaysia
2019
|
Online Access: | http://journalarticle.ukm.my/15549/1/7_ms0300_pdf_10498.pdf http://journalarticle.ukm.my/15549/ https://www.medicineandhealthukm.com/toc/14/2 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Kebangsaan Malaysia |
Language: | English |
Summary: | Point-of-care testing (POCT) of cardiac troponin device is aimed for improvement in turn round time (TAT) and assist in acute management care of acute coronary
syndrome (ACS). The present study was conducted to assess the analytical
performance and correlation of HUBI-QUANPro troponin I with an existing
laboratory instrument of high-sensitivity troponin I, Abbott Architect. The factors
that were studied, included precision study by using manufacturer quality control
(QC) material (2 levels) and correlation study of sample differences (whole blood,
plasma and serum) and methodology (immunochromatographic assay and
chemiluminescent immunoassay). A total of 30 QC was used for precision study
and 42 sample serum and EDTA for the correlation study. An acceptable total
imprecision of 10.9% and 6.7% were seen at level of 0.91 ng/mL and 2.66 ng/mL,
respectively. Regression analysis of sample differences (plasma vs whole blood)
in HUBI-QUANPro showed slope of 0.935, r=0.991 (p=<0.001). Correlation of
HUBI-QUANPro and Abbott Architect (whole blood, plasma vs serum) both
demonstrated regression slope of 0.205, r=0.963 (whole blood) and slope of 0.192,
r=0.954 (plasma), p=<0.001, respectively. HUBI-QUANPro troponin I POCT device
is a sensitive, fast, precise and has a good comparable analytical performance with
reference laboratory instrument for cardiac troponin I measurement. It is able to
serve as a good POCT device in cardiac-related acute care management. |
---|