The effect of spiral inducing chamber to left ventricular assist device outflow cannula

The rise of heart failure in the setting of shortage heart donor had brought to the emergence of development of left ventricular assist device (LVAD). However, patients are still encountering severe complications such as stroke and device thrombosis due to the hemodynamic changes. This study aimed t...

Full description

Saved in:
Bibliographic Details
Main Authors: Thomas, Siti Aisyah, Md. Khudzari, Ahmad Zahran, Abdul Kadir, Muhammad Rashidi
Format: Article
Published: Penerbit UTHM 2022
Subjects:
Online Access:http://eprints.utm.my/id/eprint/101327/
http://dx.doi.org/10.30880/ijie.2022.14.02.009
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Universiti Teknologi Malaysia
Description
Summary:The rise of heart failure in the setting of shortage heart donor had brought to the emergence of development of left ventricular assist device (LVAD). However, patients are still encountering severe complications such as stroke and device thrombosis due to the hemodynamic changes. This study aimed to compare the hemodynamic characteristic between the standard and spiral LVAD outflow cannula. Two different anastomosis sites (2cm and 3cm) from sino-tubular junction are used and a spiral inducing chamber is attached to the cannula to promote spiral flow. The result showed that spiral LVAD cannula exhibit a lower output velocity with 37% percentage difference and lower region of low WSS value.