The effect of spiral inducing chamber to left ventricular assist device outflow cannula
The rise of heart failure in the setting of shortage heart donor had brought to the emergence of development of left ventricular assist device (LVAD). However, patients are still encountering severe complications such as stroke and device thrombosis due to the hemodynamic changes. This study aimed t...
Saved in:
Main Authors: | , , |
---|---|
Format: | Article |
Published: |
Penerbit UTHM
2022
|
Subjects: | |
Online Access: | http://eprints.utm.my/id/eprint/101327/ http://dx.doi.org/10.30880/ijie.2022.14.02.009 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Universiti Teknologi Malaysia |
id |
my.utm.101327 |
---|---|
record_format |
eprints |
spelling |
my.utm.1013272023-06-08T09:03:14Z http://eprints.utm.my/id/eprint/101327/ The effect of spiral inducing chamber to left ventricular assist device outflow cannula Thomas, Siti Aisyah Md. Khudzari, Ahmad Zahran Abdul Kadir, Muhammad Rashidi Q Science (General) The rise of heart failure in the setting of shortage heart donor had brought to the emergence of development of left ventricular assist device (LVAD). However, patients are still encountering severe complications such as stroke and device thrombosis due to the hemodynamic changes. This study aimed to compare the hemodynamic characteristic between the standard and spiral LVAD outflow cannula. Two different anastomosis sites (2cm and 3cm) from sino-tubular junction are used and a spiral inducing chamber is attached to the cannula to promote spiral flow. The result showed that spiral LVAD cannula exhibit a lower output velocity with 37% percentage difference and lower region of low WSS value. Penerbit UTHM 2022 Article PeerReviewed Thomas, Siti Aisyah and Md. Khudzari, Ahmad Zahran and Abdul Kadir, Muhammad Rashidi (2022) The effect of spiral inducing chamber to left ventricular assist device outflow cannula. International Journal of Integrated Engineering, 14 (2). pp. 62-67. ISSN 2229-838X http://dx.doi.org/10.30880/ijie.2022.14.02.009 DOI: 10.30880/ijie.2022.14.02.009 |
institution |
Universiti Teknologi Malaysia |
building |
UTM Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Teknologi Malaysia |
content_source |
UTM Institutional Repository |
url_provider |
http://eprints.utm.my/ |
topic |
Q Science (General) |
spellingShingle |
Q Science (General) Thomas, Siti Aisyah Md. Khudzari, Ahmad Zahran Abdul Kadir, Muhammad Rashidi The effect of spiral inducing chamber to left ventricular assist device outflow cannula |
description |
The rise of heart failure in the setting of shortage heart donor had brought to the emergence of development of left ventricular assist device (LVAD). However, patients are still encountering severe complications such as stroke and device thrombosis due to the hemodynamic changes. This study aimed to compare the hemodynamic characteristic between the standard and spiral LVAD outflow cannula. Two different anastomosis sites (2cm and 3cm) from sino-tubular junction are used and a spiral inducing chamber is attached to the cannula to promote spiral flow. The result showed that spiral LVAD cannula exhibit a lower output velocity with 37% percentage difference and lower region of low WSS value. |
format |
Article |
author |
Thomas, Siti Aisyah Md. Khudzari, Ahmad Zahran Abdul Kadir, Muhammad Rashidi |
author_facet |
Thomas, Siti Aisyah Md. Khudzari, Ahmad Zahran Abdul Kadir, Muhammad Rashidi |
author_sort |
Thomas, Siti Aisyah |
title |
The effect of spiral inducing chamber to left ventricular assist device outflow cannula |
title_short |
The effect of spiral inducing chamber to left ventricular assist device outflow cannula |
title_full |
The effect of spiral inducing chamber to left ventricular assist device outflow cannula |
title_fullStr |
The effect of spiral inducing chamber to left ventricular assist device outflow cannula |
title_full_unstemmed |
The effect of spiral inducing chamber to left ventricular assist device outflow cannula |
title_sort |
effect of spiral inducing chamber to left ventricular assist device outflow cannula |
publisher |
Penerbit UTHM |
publishDate |
2022 |
url |
http://eprints.utm.my/id/eprint/101327/ http://dx.doi.org/10.30880/ijie.2022.14.02.009 |
_version_ |