Two-dimensional polystyrene microsphere tracking by back focal plane detection

A quadrant photodiode positioned at the back-focal plane of a microscope is used for the lateral (two-dimensional) position of polystyrene microsphere under a microscope glass slide. The computer software and hardware interface Science Workshop 750 supplied by PASCO, Inc. were used in gathering and...

Full description

Saved in:
Bibliographic Details
Main Authors: Siton, C. S., Yamson, J. M., Ayen, A. T., Pobre, Romeric F.
Format: text
Published: Animo Repository 2004
Subjects:
Online Access:https://animorepository.dlsu.edu.ph/faculty_research/12375
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: De La Salle University
Description
Summary:A quadrant photodiode positioned at the back-focal plane of a microscope is used for the lateral (two-dimensional) position of polystyrene microsphere under a microscope glass slide. The computer software and hardware interface Science Workshop 750 supplied by PASCO, Inc. were used in gathering and evaluating data yielding calibration curves and micrometer measurements of actual position displacements of the particle. Resulting SNR values computed from the experiment range from 10 to 425. Furthermore, system resolutions for both x and y lateral positions were 1.051 volts/µm and 0.915-volts/µm respectively.