A novel sensor for measuring the inner pressure of catheters for clinical use

In numerous clinical procedures requiring the infusion of a fluid (gas or liquid) it is necessary to monitor the infusion pressure. To this purpose, several devices have been developed so far. Such devices are generally disposable, with a pressure transducer in direct contact with the infusion fluid...

Full description

Saved in:
Bibliographic Details
Main Authors: Accoto, Dino, Rossini, Marco, Valentini, Simona, Portaccio, Iacopo
Other Authors: School of Mechanical and Aerospace Engineering
Format: Article
Language:English
Published: 2019
Subjects:
Online Access:https://hdl.handle.net/10356/90233
http://hdl.handle.net/10220/48535
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Nanyang Technological University
Language: English
Description
Summary:In numerous clinical procedures requiring the infusion of a fluid (gas or liquid) it is necessary to monitor the infusion pressure. To this purpose, several devices have been developed so far. Such devices are generally disposable, with a pressure transducer in direct contact with the infusion fluid. In this paper, we describe the design and development of a dry infusion pressure measurement system, sensing the pressure-induced radial displacement of the external surface of polymeric catheters. Such displacement is first amplified using a near-to-singularity elastic beam and then transduced using a Hall effect displacement sensor. We start describing the working principle of the whole system, then we detail the design process. Finally, we present the static calibration on two widely used biomedical catheters.