Performance of a portable near infrared instrument for Brix value determination of intact mango fruit
The performance of a commercially-available, portable NIR instrument "FT20" for fruit quality evaluation was investigated using mango fruits. The calibration result for the FT20 was compared with the result for the Foss NIRSystems 6500. Partial least squares (PLS) regression was used to ma...
Saved in:
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
2014
|
Online Access: | http://www.scopus.com/inward/record.url?eid=2-s2.0-0041376565&partnerID=40&md5=06425e7218619902a6218fa8ca0a398c http://cmuir.cmu.ac.th/handle/6653943832/5906 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Chiang Mai University |
Language: | English |
Summary: | The performance of a commercially-available, portable NIR instrument "FT20" for fruit quality evaluation was investigated using mango fruits. The calibration result for the FT20 was compared with the result for the Foss NIRSystems 6500. Partial least squares (PLS) regression was used to make the calibration equations. It was found that the calibration equation using spectral data obtained with the FT20 had similar accuracy to that using Foss NIRSystems 6500 spectra. The SEPs of calibration equations were 0.40°Brix for both instruments. |
---|