Feasibility of performing esophageal endoscopic submucosal dissection using master and slave transluminal endoscopic robot
The feasibility of endoscopic submucosal dissection (ESD) using the master and slave transluminal endoscopic robot (MASTER) has been shown in our previous studies [1 - 3]. Compared with gastric ESD, esophageal ESD remains challenging because of technical difficulties and risks [4, 5]. The aim of thi...
Saved in:
Main Authors: | , , , , |
---|---|
Other Authors: | |
Format: | Article |
Language: | English |
Published: |
2020
|
Subjects: | |
Online Access: | https://hdl.handle.net/10356/138023 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Institution: | Nanyang Technological University |
Language: | English |
Summary: | The feasibility of endoscopic submucosal dissection (ESD) using the master and slave transluminal endoscopic robot (MASTER) has been shown in our previous studies [1 - 3]. Compared with gastric ESD, esophageal ESD remains challenging because of technical difficulties and risks [4, 5]. The aim of this study was to evaluate the feasibility of using MASTER to perform esophageal ESD.
The new version of MASTER was used for esophageal ESD on one pig. The MASTER was redesigned to facilitate performance of ESD within a narrow working space. The main outcomes were: operating time, completeness of resection, and adverse events. The secondary outcomes included: clearance of operative field and limitation of robot arm manipulation in the narrow working space, assessed by counting the number of episodes of blind cutting. For the purpose of comparison with esophageal ESD, one gastric and one colonic ESD were performed by the same operator.
All procedures were successfully completed ([Table 1]; [Figs. 1 – 3]; [Video 1]). It took 15, 63, and 45 minutes to complete the esophageal, gastric, and colonic ESDs, respectively. All lesions were excised en bloc; the specimen sizes were: 20 × 20, 50 × 45, and 35 × 35 mm, respectively. The dissection speeds were: 20.9, 28.0, and 21.4 mm2/min, respectively. There were no adverse events. The number of episodes of blind cutting were: 2, 6, and 6 (0.13, 0.10, and 0.13/min). In contrast to gastric and colonic ESD, the traction of the grasper during esophageal ESD tended to align with the long axis because of the narrow working space. There was no difference comparing esophageal against gastric and colonic ESD in terms of the speed of dissection, the rate of complete resection, the occurrence of adverse events, and the number of episodes of blind cutting. In conclusion, performing esophageal ESD using the MASTER was feasible with a certain degree of adjustment for the narrow working space. |
---|