The influence of chromatic aberration on the dose-limited spatial resolution of transmission electron microscopy

The effect of chromatic aberration (CC) on the spatial resolution in transmission electron microscopy (TEM) was studied in thick specimens in which the sample becomes the limiting factor in the resolution. The sample influences the energy spread of the electron beam, allows only a limited electron d...

Full description

Saved in:
Bibliographic Details
Main Authors: Ortega, Eduardo, Boothroyd, Chris, de Jonge, Niels
Other Authors: School of Materials Science and Engineering
Format: Article
Language:English
Published: 2023
Subjects:
Online Access:https://hdl.handle.net/10356/165539
Tags: Add Tag
No Tags, Be the first to tag this record!
Institution: Nanyang Technological University
Language: English
Description
Summary:The effect of chromatic aberration (CC) on the spatial resolution in transmission electron microscopy (TEM) was studied in thick specimens in which the sample becomes the limiting factor in the resolution. The sample influences the energy spread of the electron beam, allows only a limited electron dose, and modulates electron scattering events. The experimental set-up consisted of a thin silicon nitride membrane and a silicon wedge containing gold nanoparticles. The resolution was measured as a function of electron dose and sample thickness for different sample configurations and for different microscopy modalities including regular TEM, energy filtered TEM (EFTEM) and CC-corrected TEM. Comparison with an analytical model aided the understanding of the experimental data applied over varied conditions. The general trend for all microscopy modalities was a transition from a noise-limited resolution at low electron dose to a CC-limited resolution at high-dose in the absence of beam blurring. EFTEM required an accurate energy slit offset and an optimal energy spread to energy-slit width ratio to surpass regular TEM. The key advantage of CC correction appeared to be the best possible resolution for larger sample thickness at low electron dose outperforming EFTEM by about fifty percent. Several hypothetical sample configurations relevant to liquid phase electron microscopy were evaluated as well to demonstrate the capabilities of the analytical model and to determine the most optimal microscopy modality for this type of experiment. The analytical model included an automated optimization of the EFTEM settings and may aid in optimizing the sample-limited resolution for experimental analysis and planning.